Desktop multichannel signal generation calibrator
Products overview:
ZW - Y200B seriesDesktop multipath signal checkInstrument that can simulate all kinds of industrial process of sensor and transmitter, can lose 11 standard signals at the same time, also can measure some routine instrument produced by all kinds of signals, so is suitable for a new generation of automation instrument, automatic control system commissioning and validation.Equipped with 24 VDC output for adjustable two wire measurement instrument, so use the ZW - 20 b desktop multipath signal after verification without having to match other instruments can be easily and accurately for more than one instrument for debugging and testing at the same time.
Second, the widely used:
ZW - Y200B seriesDesktop multipath signal checkIs a kind of intelligent instrument verification, high precision, small volume, can be used in the field of industrial instrumentation calibration, also can be used in the calibration of laboratory instruments and meters.Powerful, can simulate the output to a variety of industrial control in the process of measurement and control signals, and at the same time also can measure these occurring in the process of industrial control signal.
Three,Desktop multipath signal occurred verification key features:
1, provide DC24V circuit of power supply, directly adjusted two wire measurement instrument;
2, 5 LCD display;
3, high precision, 0.05%;
4, automatic range, can effectively improve the precision;
5, the built-in battery, a charge can work 20 hours (Ni MH battery);
6, frequency measurement resolution can reach 0.01 Hz;
7, 20.000 mA, such as 5.0000 V display five readings.
Four,Desktop multipath signal occurred verification technology parameters:
1, the output current: maximum 20 ma, 4-way independence, accuracy of 0.05 on the Richter scale;
2, the load resistance: RL 1500 or less Ω (0 to 10 ma) RL 750 or less Ω (0 to 20 ma);
3, millivolt output: 120 ma, the largest precision of 0.05 on the Richter scale;
4, load resistance: RL Ω 10 k or higher voltage output: maximum 6 v, accuracy of 0.05 on the Richter scale;
5, frequency output: F1, F2 10000 hz, the largest maximum error 2 hz;
6, the biggest: material 15 v F3, F4 maximum 1000.0 Hz, 0.2 Hz maximum error;
7, resistance output: 0-390 Ω current measurements: 0 to 20 ma, Ri = 100 Ω, accuracy of 0.05 on the Richter scale;
8, millivolt measurement: 0-200 mv, Ri = 10 m Ω, accuracy of 0.05 on the Richter scale;
9, voltage measurements: 0 to 20 v, Ri = 10 m Ω, accuracy of 0.05 on the Richter scale;
10, frequency measurement: 0-20 KHZ;
11, sensitivity: the Vp - p p 500 mv,
12, accuracy: 0.02% plus or minus 1 word;
13, the display mode: 4 1/2 a super bright LED display;
14, using temperature: - 10 ℃ to 50 ℃
15, environment humidity: 95% relative humidity, or less without condensation;
16, electricity source, ac 220 v + 10%, 50 hz.
Five,Measurements:
range | The maximum | precision |
500.00mV | 60V | 0.05%+3d |
5.0000V | 0.05%+3d | |
50.000V | 0.05%+3d | |
20.000mV | 240mV | 0.05%+3d |
200.00mV | 0.05%+2d | |
20.000mA | 240mA | 0.05%+2d |
200.00mA | 0.05%+2d | |
0-500KHz Five automatic range | 650KHz | 0.15%+2d |
Vi.Technical indicators:
range | The maximum | precision |
500.00mV | 60V | 0.05%+3d |
5.0000V | 0.05%+3d | |
200.00mV | 240mV | 0.05%+5d |
20.000mA | 24mA | 0.05%+2d |
2.0000KHz | 22KHz | 0.05%+3d |
20.000KHz | 0.05%+3d |
Seven warranty and after-sales service:
ZW - Y200B series desktop multipath signal check product quality assurance period for 12 months, life-long maintenance.Warranty period for the damage, free maintenance, more than warranty, maintenance fee and shipping fee only.To new users first use, generally by the menu function prompts the operation, there are technical difficulties, please call the company technical department to give answers. Company can send engineers to the scene to guide if necessary.