国产精品视频久久久久久|亚洲精品视频网站在线观看视频|国产极品白嫩美女在线观看看|欧美一区和二区视频在线观看|国产美女在线观看精品网站免费

Hello! Welcome to Jiangsu Zhongwei Measurement and Control Instrument Co., Ltd.
0517-86918678
Address:No. 4-4 East West Road, Jinhu County Industrial Concentration Zone, Huaian City, Jiangsu Province
Tel:0517-86918678
Fax:0517-86918679
Email:zhongweick@163.com
Post Encoding:211600

EJA118W Diaphragm Sealed Differential Pressure Transmitter

EJA118W隔膜密封差壓變送器.jpg
  • EJA118W隔膜密封差壓變送器.jpg
EJA118W diaphragm sealed differential pressure transmitter uses ASIC amplifier design to reduce pack…

Description:

A small, lightweight design:
Diaphragm sealed differential pressure transmitterThe ASIC amplifier design to make the packaging becomes small, and the miniaturization of the membrane structure and the flange of the box, this type of weight dropped to half of the original type.ASIC design not only reduces the number of parts, but also improve the reliability of the amplifier.
Two, field bus communication ability:
Fieldbus is a two-way digital communication system, is the innovative technology configuration of the instrument control system, but also the prospect of upgrading good product updates, for the replacement of standard 4~20mA is widely used in analog communication instrument.
The EJA series provides two kinds of fieldbus, FOUNDATION fieldbus? Low voltage mode and PROFIBUS PA devices, to realize the collaborative work of Henghe motor and other suppliers.The software, EJA series uses two AI function block calculation of differential pressure and static pressure, so as to realize flexible instrument configuration.
Three, main features:
The world's first - silicon resonant sensor
Silicon resonant sensor.
High accuracy: + 0.075%;
5 years of continuous work without adjustment;
The temperature effect is negligible;
The static effects can be ignored.
Effect: 100 thousand consecutive unidirectional voltage unidirectional overvoltage after the experiment affect the amount of less than 0.03%;
The function of two-way communication (BRAIN/HART protocol, FF bus).
Self diagnosis function perfect;
A small, light weight (standard 3.9kg);
The static effect is negligible:
When a static pressure (pressure), resonant beam deformation two shape, size and material identical to the same, so the frequency change is consistent, so the deviation automatically cleared (figure similar formula and temperature influence).
Excellent characteristics of unidirectional overvoltage:
The liquid membrane and the membrane using corrugated box body processing technology of the original, so that the external pressure increases to a certain value, the film can fully contact with the body, silicone oil is transferred to the sensor pressure no longer increase with the increase of external force, so as to achieve the protective effect on sensor.
Four, standard specification:
1, communication line:
Supply voltage: 16.4 ~ 42V DC (16.4 30V DC);
Communication distance: 2km, use the CEV cable;
The load capacitance: 0.22 F;
Load inductance: below 3.3mH;
Space and power line: more than 15cm;
The receiving instrument is connected to the receiving of resistance on the input impedance in 2.4kHz, 10K. The above ambient temperature -40 ~ 85 (-40 ~ 185 F) (general); -30 (-22 ~ 80 degrees to 176 degrees (F) with digital meter).
2, working pressure: 7kPaabs (20mmHgabs) to the flange of the rated working pressure.
Effect of 0.005%/V 3, the power supply voltage variation (21.6 ~ 32VDC, 350).
4, installation:
Transmitter: 2-inch pipeline installation;
Seals: flange installation.
5, the installation of flange specifications:
A flange and gasket ANSI specification for contact, zigzag machining (according to ANSIB16.5)
6, waterproof structure:
JIS C0920 anti immersion type (equivalent to IECIP67 and NEMA4X)
7 transmitter material:
Fran: JISSUS14A
Flange bolt: refer to the "type and specification code table"
8, diaphragm seal material:
Capillary: JIS SUS316;
Protective tube: JIS SUS304;
PVC (Zui shield PVC use temperature: 100 degrees).
9, the damping time constant:
Time constant sum amplifier assembly and diaphragm sealing film box assembly value;
The amplifier assembly time constant: 0.2 ~ 64 seconds, can be divided into 9 sections set;
When the capillary length is 3M, the injection of code A, obtained under the normal temperature value.

Wu, specifications range:

application

type

Model

Film box

Range (KPa)

Zui working pressure

Differential pressure and liquid level (diaphragm sealed)

Convex diaphragm

EJA118N

EJA118W

EJA118Y

M

2.5-100

Based on the flange specifications

Flat diaphragm

A flat

A convex

H

25-500

Six, product selection list:

Basic specification code:EJA118Y-DMSJ12PA-AC01-92NN

First: output signal

-D4-20mA DC/BRAIN -E4-20mA digital communication protocol DC/HART protocol digital communication -FFF fieldbus communication

 

Second: the measurement range (film box)

M2.5-100KPa (250-1000mmH2O) H25-500KPa (0.25-5kgf/cm2)

 

Third: high side (convex shaped diaphragm) with a liquid material

S[]JIS SUS316L]JIS SUS316 tube diaphragm [[]JIS SUS316

 

Fourth: flange specifications

J1JIS 10KJ2JIS 20KA1ANSI 150A2ANSI 300P1JIP 150P2JIP 300D2DIN PN10/16D4DIN PN25/40G2GB PN10/16G4GB PN25/40G6GB PN100

 

Fifth: diaphragm protruding length (X2)

2X2=150mm4X2=100mm6X2=150mm

 

Sixth: flange size / material

P high side 100mm (4 ") JIS S25C, 80mm low voltage side (3") /JIS S25CQ 100mm high side (4 ") JIS SUS304, 80mm low voltage side (3") /JIS SUS304R 100mm high side (4 ") JIS SUS316, 80mm low voltage side (3") /JIS SUS316

 

Seventh: bolt and nut material

AJIS SCM435BJIS SUS 630

 

Eighth: filling liquid

-A general use (silicone oil) [liquid contact temperature]-10-250 [environmental temperature]-10-60 -B general use (silicone oil) [liquid contact temperature]-30-180 [environmental temperature]-15-60 -C with high temperature (silicone oil) [liquid contact temperature]10-300 [environmental temperature]10-60 -D with no oil (fluorine oil) [] with a liquid temperature -20-120 [C environment temperature]-10-60 -E with low temperature (ethylene glycol) [liquid contact temperature]-50-100 [environmental temperature]-40-60

 

Ninth: -

C: high side pressure flange behind, low voltage side: side pressure.

 

Tenth: the capillary length (m)

01 1m02 long 2m03 long capillary capillary capillary capillary 3m04 long long 4m05 long 5m06 long 6m07 capillary capillary capillary capillary long 7m08 long 8m09 long 9m10 long capillary capillary 10m

 

Eleventh: installation

The level of -9 piping connection, the left front leads to high pressure, capillary

 

Twelfth: electrical interface

0G1/2 thread, a thread of 21/2NPT electrical interface, two electrical interface, no blind plug 3Pg13.5 internal thread, two electrical interface, no blind plug 4M20 internal thread, two electrical interface, no blind plug 5G1/2 internal thread, two electrical interface, a blind plug 71/2NPT internal thread, two electrical interface, a blind plug 8Pg13.5 internal thread, two electrical interface, a blind plug 9M20 internal thread, two electrical interface, a blind plug

 

Thirteenth: built-in display table

D E digital display table with a measurement range setting button digital display meter N

 

Fourteenth: mounting bracket

AJIS SECC 2 "pipe installation (Pinto shape) BJIS SUS304 2" pipe installation (Pinto shape) N


Browse:
Return To The Top

分享到: